Temperature data logger for autoclaves resistant up to 150°C

MicroWH L is a temperature data logger from -40°C to 150°C (standard calibration from 25°C to 150°C) with 20 mm or on demand length external probe (probes cannot be switched) on a 3 mm base (base height is not counted for probe length), managed with Windows software and USB interface (DiskInterface HS, Multibay). Battery is user replaceable and the data logger is provided with a National Reference traceable cetificate on 6 points.

Note: if used below -30°C, battery life will be highly reduced.

Video

Applications

Main features

  • With different lengths rigid probe for penetration
  • Completely food grade and waterproof
  • All software calculate lethality value (F0, PU, A0 ecc.)
  • Low battery consumption for an extended battery life
  • User replaceable battery (software shows battery status)
  • Very easy to deploy in any type of package
  • National Reference traceable calibration certificate included
  • For calibration over 140°C the minimum probe length is 50 mm

Plus

  • Extremely high accuracy and precision: with an accuracy of ± 0,1°C these devices can be employed in any application involving pharmaceuticals, validation, laboratory and medical field
  • High accuracy even outside the calibration range
  • Fast response time thanks to the 3 mm diameter probe
  • Printed reports compliant with health regulations and ISO (data are not editable in the software)

The system

The system is made up by:

  • MicroWH L temperature data logger
  • DiskInterface HS or Universal Multibay
  • SPD software or TS Manager software (compatible with the FDA 21 CFR Part 11 regulation)

 

Dimensions 39 h X 20 Ø (mm)
Probe dimensions Probe base dimensions 3 h X 14 Ø (mm) - Probe 20/on demand l X 3 Ø (mm) (l on demand: min. 12 mm / max. 175 mm. For longer probes ask for quotation)
Weight 50 g
Materials Stainless steel AISI316L, PEEK
Temperature range -40°C ÷ +150°C
Sensor temperature range -40 °C ÷ +250 °C
Standard calibration points (temperature) 25/50/75/100/121/150°C
Extra calibration points (temperature) Within the range -40 °C ÷ +150 °C
Temperature resolution 0,01 °C
Temperature accuracy ± 0,1 °C (within the calibration range) / ± 0,05 °C with ultra accuracy option
Memory (n. of acquisitions) 20.224
Acquisition step From 1 every second up, with 1 second steps
Protection degree IP68
Battery life +10.000.000 acquisitions at 1 second step continuously (calculated time @25°C. Battery life may be shorter if used in low temperatures)
Software&Mobile App SPD, TS Manager
Accessories DiskInterface HS, Multibay universale