RHTemp80 L
RHTemp80 L is a temperature and humidity data logger from -40°C to 80°C managed with Windows software and USB interface (DiskInterface HS, Multibay). …
Rotronic Hygroclip temperature and humidity intelligent sensors with high accuracy, no cable; for longer cable length purchase ZED extensions.
Intelligent sensors are much more than simple sensors. The idea of an intelligent sensor is based on the fact that its parameters (serial number, type etc.) and the calibration data are stored electronically in a circuit installed on its cable or inside the connector. In this way you can connect to a reading device or directly to your PC, several intelligent sensors of different kind and they will all be recognized for what they are and what they are able to read, along with their calibration data. The sensor sends to the device (PC or data logger) a digital calibrated data.
The main feature of the intelligent sensors, used in the Syrinx system and connected to the ZED modules, is the recalibration procedure: easy, quick and not expensive. You won't need to remove the ZEDs from their position. You won't need to send the sensors to Tecnosoft for recalibration and therefore remain without monitoring. Before calibration expires order the recalibration service for intelligent sensors. You will receive new sensors to replace the expiring ones that will be sent back to Tecnosoft later on.
All intelligent sensors come with an Accredia (NIST equivalent) traceable calibration certificate included.
Intelligent sensors work with the Syrinx system and the FridgeLog Z software:
Intelligent sensor dimensions | 83 h X 15 Ø (mm) |
---|---|
Sensor temperature range | -40 °C ÷ +80 °C |
Standard calibration points (temperature) | 5/30/50°C |
Extra calibration points (temperature) | Within the range -40 °C ÷ +80 °C |
Temperature resolution | 0,05 °C |
Temperature accuracy | ± 0,3 °C (within the calibration range) |
Humidity range | 0% ÷ 100% RH (not condensing) |
Standard calibration points (humidity) | 30/50/70% @ 30°C ± 10°C |
Extra calibration points (humidity) | Within the range 10% ÷ 90% RH |
Humidity reslution | 0,1% |
Humidity accuracy | ± 2% RH |