Subscribe to our newsletter - Tech support
1989 / 2024 Flip through our history

USB 8 channels temperature recorder for K type thermocouples thermocouples

Temperature data loggers for thermocouples up to 8 channels, resistant up to 130°C, with miniature connectors for K type thermocouples. Operative range is from -60°C to 130°C.

User replaceable battery. Does not need of any connection bay, it has a direct USB connection. Thermocouples not included.

 

Download

TC-Log 8 USB (all versions) - USB Drivers

Software download and related accessories

Applications

Main features

  • High temperatures resistant
  • Quick response time
  • USB direct connection to PC
  • Reduced size
  • Real time monitoring if connected to a PC via the USB interface
  • User calibrable

 

Plus

  • Easy to use
  • You can use your own thermocouples and calibrate it with them
  • Real time

The system

System components:

  • TC-Log 8 USB K
  • Thermocouples
  • TC-Log software

 

Dimensions 132 h X 80 X 25 (mm)
Weight 190 g
Temperature range -60°C ÷ +130°C
Number of channels 8
Materials Aluminium
Mounted connectors K
Available thermocouples K
Temperature resolution 0,15 °C
Temperature accuracy For logger only, sensors excluded: ± 0,2 °C with Accredia traceable certificate (calibration at -40/25/125°C)
Temperature accuracy For logger only, sensors excluded: ± 0,2 °C with Accredia traceable certificate (calibration at room temperature)
Temperature accuracy For logger only, sensors excluded: +/- 0,5°C range 0°C ÷ 75°C; +/-1°C range -40°C ÷ 0°C and 75°C ÷ 130°C without calibration
Stability time Internal stability after a 50°C shock: 20 minutes
Memory (n. of acquisitions) 104.855 with only 1 channel, 16.383 with all 8 channels (for complete information check proper document)
Acquisition step From 1 every second up, with 1 second steps
Battery life +4.320.000 acquisitions at 1 second step continuously (calculated time @25°C. Battery life may be shorter if used in low temperatures)
Battery life Infinite time at 1 second step continuously when connected to USB port, real time (calculated time @25°C. Battery life may be shorter if used in low temperatures)
Battery type AA 3.6V Lithium Thionyl Chloride, 2Ah
Software&Mobile App TC-Log

Documentation

Faq