MicroW L Flexible
MicroW L Flexible is a temperature data logger from -40°C to 140°C (calibration from 25°C to 140°C) with 30 cm felxible probe and 50 mm or on …
MicroW L is a temperature data logger from -40°C to 140°C with 20 or 50 or 100 or 150 mm or on demand length external probe (probes cannot be switched) on a 3 mm base (base height is not counted for probe length), managed with Windows software and USB interface (DiskInterface HS, Multibay). Battery is user replaceable and the data logger is provided with an Accredia (NIST equivalent) traceable cetificate on 6 points (standard calibration from 25°C to 140°C, calibrable from -40°C to 250°C, only the probe will resist above 140°C. For calibrations above 140°C probe minimum length is 50 mm).
Note: if used below -30°C, battery life will be highly reduced.
The system is made up by:
Dimensions | 39 h X 20 Ø (mm) |
---|---|
Probe dimensions | Probe base dimensions 3 h X 14 Ø (mm) - Probe 20/50/100/150/on demand l X 3 Ø (mm) (l on demand: min. 12 mm / max. 175 mm. For longer probes ask for quotation) |
Weight | 50 g |
Materials | Stainless steel AISI316L, PEEK |
Temperature range | -40°C ÷ +140°C |
Sensor temperature range | -40 °C ÷ +250 °C |
Standard calibration points (temperature) | 25/50/75/100/121/140°C |
Extra calibration points (temperature) | Within the range -40 °C ÷ +250 °C |
Temperature resolution | 0,01 °C |
Temperature accuracy | ± 0,1 °C (within the calibration range) |
Memory (n. of acquisitions) | 20.224 |
Acquisition step | From 1 every second up, with 1 second steps |
Protection degree | IP68 |
Battery life | +10.000.000 acquisitions at 1 second step continuously (calculated time @25°C. Battery life may be shorter if used in low temperatures) |
Software&Mobile App | SPD, TS Manager |
Accessories | DiskInterface HS, Multibay universale |